Flow
...
Controller
Linearization and Calibration
Introduction
Presently, the flow control module is being utilized to create a constant head tank, where the water level in a tank remains constant, being The Flow Controller device is used to maintain a constant level of water in a source, a constant head tank, which is refilled by a stock tank. This By adjusting the constant head tank is the measuring point for the head loss which governs the flow rate of Alum into the plant. However when implemented in the field, it has become increasingly apparent that there are one or more, possibly yet unidentified, confounding variables that are rendering our equations inaccurate, causing for , different flow rates of Alum may be sent to the plant for the flocculation process. Field use of this device, however, is consistently inaccurate, diverging in direction relation to flow rate as well as tubing length. However, the majority of the data seems to indicate that there are in fact more confounding variables which make flow predictions nearly impossible. This dependence on faulty information, namely the inapplicable equations, may cause miss-dosing in the plant , reducing efficiency and at times creating wasteful and dangerous situations for the communities.or a wasteful situation of using too much Alum for the plant.
Divergence specifically occurs to cause under delivery of alum to the plant, which decreases the efficacy of flocculation and has an adverse affect on effluent turbiditySpecifically, in tests that have been conducted in Honduras and the United States, flow rates at greater head losses are expirimentally producing lower flow rates that those calculated, indidcating that the flow control module is not functioning linearly. This section of the Flow Control Module Team is conducting tests to identify confounding variables to more accurately render parameterize the equations to provide for more efficient flocculation in the plant.
Goals
This project section is attempting to resolve the mystery of where inconsistencies arise that cause deviations from the standardized equations that relate head loss and flow rates which are used to dose the flow of the water treatment plant. There are a few variables that we hope to test. First we need to assure our selves that the flow control module is actually operating within the laminar flow range, rather than prematurely transitioning to the turbulent flow range. After this, we will look into the affect of tube length on the flow, as well as evaluating the possible affect that the length of the tube in question has on the dosing flow that is being modeled. After this, the question of additional sources of head loss will be considered, the possiblity of manufacturing in acuracies or possible the inclusion of constants based on the mechanism. Lastly, we will look into the possiblity that highly concentrated solutions of alum are causing unintending consequences that need to be factored inteam is working specifically with the constant head tank as a distribution system with known parameters and an adjustable arm for dosing providing a known head loss from elevation. First it is important to verify that the tubes in use, do in fact, meet the specifications of smoothness and diameter. Once we are sure that the material is regular, and if not, once we may correct for this difference, it is important to see what the trend is in divergence from the Hagen-Poiseuille equation. Graphing of flow rate vs. Head loss allows for a simply relationship between the experimental and theoretical. Once these have been parameterized for the purposes of calculation, it will be useful to continue looking into possible sources of head loss through the system or any factors that may deter flow.
These goals will be realized by the utilization of the depicted apparatus. The constant head tank was constructed using two elevated buckets. The storage was significantly larger and with a larger diameter than the actual constant head bottle. They were connected by a single tube which went from the base of the stock tank and into the top portion of the bottle where the float was located. From the bottom of the bottle was a from which the dosing tube was attached. This tube is adjustable to allow for different lengths. A PVC pipe was procured and sealed on the bottom and placed in a wooden base to stabilize it. In this, along the top 26 centimeters, 26 7 mm holes were drilled. The top hole had its center line leveled with the water level in the constant head bottle so that ever additional hole below was another centimeter of head loss.
Additional Documentation
The Linearization and Calibration Quiz- find out how much you really know about this flow controller sub team.
Linearization and Calibration team goals and meeting minutes.
Methods and Results from Specific Experiments
Flow Rate vs. Headloss for Variable Tube Lengths
- Taking the new apparatus, flow rates were calculated relative to head loss and plotted again the Hagen-Poiseuille equation to analyze divergence from the mathematical model
Verification of Tubing Inner Diameter
- Calculate the inner diameter of the black tube used for dosing to ensure accuracy in dosing
...
- Experiments were conducted to establish a model for the relationship between FCM outflow and headloss head loss associated with outflow tube position
...
- The maximum inlet pressure that the float valve can shut off was investigated in order to determine how much head there can be between the chemical stock tanks and the flow controllers
- Background on previous research
Conclusions
Based on the research that has been done, it It may be concluded that the flow control module is prematurely transitioning into turbulent flow, causing inaccuracies in the calculated flows. We hope to continue testing, modifying different variables to find the source of the inaccuracythere are definitely variables that are causing significant distortion between experimental and theoretical results. Possibly due to un-considered sources of head loss, experimental results are producing dramatically lower flow rates than are predicted theoretically. This may be attributed to manufacturing error, specifically in the diameter or roughness of the dosing tube; the corrective abilities of these terms on the equations are note worth. Additional challenges as they relate to future topics may be found here.